The perfusion index measured by the pulse oximeter affects the agreement between ClearSight and the arterial catheter-based blood pressures: A prospective observational study.


ABSTRACT:

Background

ClearSight is a noninvasive arterial blood pressure monitor, but it remains unknown whether it is affected by the state of perfusion to the fingers. We investigated whether the lower perfusion index (PI) measured with a pulse oximeter, which reflects finger perfusion, would affect the agreement between arterial pressures measured with ClearSight versus those obtained with an arterial catheter.

Methods

Paired arterial pressure data (ClearSight and arterial catheter-based pressures) and PI values were prospectively obtained from 30 patients undergoing major abdominal surgery. The primary outcome was standard deviation (SD) of the bias (precision) of blood pressure between ClearSight and arterial catheter. The ratio of the adjusted SD of the bias between PI≤1 and PI>