Unknown

Dataset Information

Measurement of the liver iron concentration in transfusional iron overload by MRI R2* and by high-transition-temperature superconducting magnetic susceptometry.


ABSTRACT:

Purpose

To compare measurement of the liver iron concentration in patients with transfusional iron overload by magnetic resonance imaging (MRI), using R2*, and by magnetic susceptometry, using a new high-transitiontemperature (high-Tc; operating at 77 K, cooled by liquid nitrogen) superconducting magnetic susceptometer.

Methods

In 28 patients with transfusional iron overload, 43 measurements of the liver iron concentration were made by both R2* and high-Tc magnetic susceptometry.

Results

Measurements of the liver iron concentration by R2* and high-Tc magnetic susceptometry were significantly correlated when comparing all patients (Pearson's r = 0.91, p < 0.0001) and those with results by susceptometry >7 mg Fe/g liver, dry weight (r = 0.93, p = 0.006). In lower ranges

SUBMITTER: Sheth S 

PROVIDER: S-EPMC6626578 | biostudies-literature | 2019 May - Jun

REPOSITORIES: biostudies-literature