Permanent and Transient Electrophysiological Effects During Cardiac Cryoablation Documented by Optical Activation Mapping and Thermal Imaging.


ABSTRACT: Cardiac catheter cryoablation is a safer alternative to radiofrequency ablation for arrhythmia treatment, but electrophysiological (EP) effects during and after freezing are not adequately characterized. The goal of this study was to determine transient and permanent temperature induced EP effects, during and after localized tissue freezing. Conduction in right (RV) and left ventricles (LV) was studied by optical activation mapping during and after cryoablation in paced, isolated Langendorff-perfused porcine hearts. Cryoablation was performed endocardially (n=4) or epicardially (n=4) by a cryoprobe cooled to -120 °C for 8 minutes. Epicardial surface temperature was imaged with an infrared camera. Viability staining was performed after ablation. Motion compensation and co-registration was p

SUBMITTER: Morley G 

PROVIDER: S-EPMC6667312 | biostudies-literature | 2018 Nov

REPOSITORIES: biostudies-literature