Ontology highlight
ABSTRACT: Introduction
Cardiac catheterization procedures result in high radiation doses and often multiple procedures are necessary for congenital heart disease patients. However, diagnostic reference levels (DRL) remain scarce. Our first goal was finding the optimal DRL parameter and determining appropriate DRLs. The second goal was to calculate organ doses (OD), effective doses (ED) and lifetime attributable risks (LAR) per procedure and to provide conversion factors based on dose area product (DAP).Materials and methods
DRLs are calculated for each procedure type, as the 75th percentile of the cumulative value per procedure from the corresponding parameter. All irradiation events in the DICOM Structured Reports were automatically processed and simulated using PCXMC, resulting in
SUBMITTER: Buytaert D
PROVIDER: S-EPMC6668795 | biostudies-literature | 2019
REPOSITORIES: biostudies-literature