Reliability of a portable device for quantifying tone and stiffness of quadriceps femoris and patellar tendon at different knee flexion angles.


ABSTRACT: The reliability of MyotonPRO that can monitor the mechanical properties of tissues is still unclear. This study aimed to analyze the within-day inter-operator and between-day intra-operator reliability of MyotonPRO for assessing tone and stiffness of quadriceps femoris and patellar tendon at different knee angles. The tone and stiffness of healthy participants (15 males and 15 females, aged 24.7±1.6 years) in the supine and resting position were measured using the MyotonPRO device. The measurements were quantified at 0°, 30°, 60°, and 90° of knee flexion. The intraclass correlation coefficient (ICC), standard error of measurement (SEM), and minimal detectable change (MDC) were calculated and a Bland-Altman analysis was conducted to estimate reliability. The results indicated excellent inte