Ontology highlight
ABSTRACT: Objective
To determine the intra-, inter- and test-retest variability of CT-based texture analysis (CTTA) metrics.Materials and methods
In this study, we conducted a series of CT imaging experiments using a texture phantom to evaluate the performance of a CTTA panel on routine abdominal imaging protocols. The phantom comprises of three different regions with various textures found in tumors. The phantom was scanned on two CT scanners viz. the Philips Brilliance 64 CT and Toshiba Aquilion Prime 160 CT scanners. The intra-scanner variability of the CTTA metrics was evaluated across imaging parameters such as slice thickness, field of view, post-reconstruction filtering, tube voltage, and tube current. For each scanner and scanning parameter combination, we evaluated the perfo
SUBMITTER: Varghese BA
PROVIDER: S-EPMC6698768 | biostudies-literature | 2019 Aug
REPOSITORIES: biostudies-literature