ABSTRACT: Objective
The objective of this study was to investigate and compare the hemodynamic and turbulence characteristics upon implantation of St Jude Medical (SJM) (St Jude Medical, St Paul, Minn) and On-X (On-X Life Technologies, Kennesaw, Ga) bileaflet mechanical valves. Both valves are considered highly successful bileaflet mechanical valves characterized by good clinical outcomes despite their numerous design differences. Although thromboembolism remains the main disadvantage of bileaflet mechanical valves, On-X valves have been shown to need less anticoagulation therapy.Methods
Hemodynamic assessment of a 23-mm On-X bileaflet mechanical valve and a 23-mm bileaflet SJM valve implanted in an aortic root was performed under pulsatile physiologic conditions. Time-resolved and p