Metal implants influence CT scan parameters leading to increased local radiation exposure: A proposal for correction techniques.
Ontology highlight
ABSTRACT: Metal implants not only deteriorate image quality, but also increase radiation exposure. The purpose of this study was to evaluate the effect of metal hip prosthesis on absorbed radiation dose and assess the efficacy of organ dose modulation (ODM) and metal artifact reduction (MAR) protocols on dose reduction. An anthropomorphic phantom was scanned with and without bilateral metal hip prostheses, and surface and deep level radiation doses were measured at the abdomen and pelvis. Finally, the absorbed radiation doses at pelvic and abdominal cavities in the reference, ODM, and two MAR scans (Gemstone spectral imaging, GE) were compared. The Mann Whitney-U test and Kruskal-Wallis test were performed to compare the volume CT dose index (CTDIvol) and mean absorbed radiation doses. Unilateral an