Influence of antibody immobilization strategy on carbon electrode immunoarrays.


ABSTRACT: We report here the influence of antibody immobilization strategy for protein immunosensors on screen printed carbon electrode arrays in terms of antibody binding activity, analytical sensitivity, limit of detection, and stability. Horseradish peroxidase (HRP) was the model analyte with anti-HRP immobilized on the sensors, and HRP activity was used for detection. Covalently immobilized anti-HRP antibodies on electrodes coated with chitosan, electrochemically reduced graphene oxide (rGO), and dense gold nanoparticle (AuNP) films had only 20-30% of the total immobilized antibodies active for binding. Active antibodies increased to 60% with passively adsorbed antibodies on bare electrodes, to 85% with oriented antibodies using protein A covalently immobilized on AuNP-coated carbon electrode, a

SUBMITTER: Sharafeldin M 

PROVIDER: S-EPMC6711173 | biostudies-literature | 2019 Sep