A Novel Bismuth-Chitosan Nanocomposite Sensor for Simultaneous Detection of Pb(II), Cd(II) and Zn(II) in Wastewater.
Ontology highlight
ABSTRACT: A novel bismuth (Bi)-biopolymer (chitosan) nanocomposite screen-printed carbon electrode was developed using a Bi and chitosan co-electrodepositing technique for detecting multiple heavy metal ions. The developed sensor was fabricated with environmentally benign materials and processes. In real wastewater, heavy metal detection was evaluated by the developed sensor using square wave anodic stripping voltammetry (SWASV). The nanocomposite sensor showed the detection limit of 0.1 ppb Zn2+, 0.1 ppb Cd2+ and 0.2 ppb Pb2+ in stock solutions. The improved sensitivity of the Bi-chitosan nanocomposite sensor over previously reported Bi nanocomposite sensors was attributed to the role of chitosan. When used for real wastewater samples collected from a mining site and soil leachate, similar detectio
SUBMITTER: Hwang JH
PROVIDER: S-EPMC6723456 | biostudies-literature | 2019 Jul
REPOSITORIES: biostudies-literature
ACCESS DATA