Performance of Dynamic Automated CT Annular Measurements Compared to Standard Manual Measurements for Transcatheter Aortic Valve Replacement Sizing.


ABSTRACT:

Purpose

We sought to determine the performance of an automated computed tomography (CT) software that provides dynamic annular measurements of all available cardiac phases for transcatheter aortic valve replacement (TAVR) sizing as compared to the standard single manual measurement.

Materials and methods

In 60 TAVR patients (84±7 years, 60% male) who underwent pre-procedural CT scans, we measured the aortic annular diameters, perimeter, and area using (1) the dynamic automated CT measurements and (2) standard single manual measurement from the cardiac phase of maximum systolic opening by visual estimate.

Results

The automated software was successful in providing annular measurements in 43/60 (72%) of cases, with the remainder requiring semi-automated contours. The max