High-volume workflow and performance comparisons for Chlamydia trachomatis and Neisseria gonorrhoeae testing using automated molecular platforms.


ABSTRACT:

Background

The global burden of sexually transmitted infections (STIs) is high and there have been reports of increasing chlamydial and gonorrheal infections. High-volume screening programs for Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (NG) are an important component of STI control. This study evaluated the high-volume workflow and performance of the cobas® CT/NG assay for use on the automated Roche cobas® 6800 system, with the cobas p 480 instrument for pre-analytics, compared with the Aptima Combo 2 assay on the Hologic Panther system.

Methods

High-volume workflow and performance were evaluated using paired female urine specimens. Workflow analysis (n = 376) included hands-on time (HoT), number of manual interventions, and time to first and last results. For per

SUBMITTER: Frontzek A 

PROVIDER: S-EPMC6737607 | biostudies-literature | 2019 Sep