Registry study on failure incidence in 1,127 revised hip implants with stem trunnion re-use after 10 years of follow-up: limited influence of an adapter sleeve.


ABSTRACT: Background and purpose - Little is known about the role of retained trunnions in revision hip arthroplasties, i.e., when only the femoral head is substituted. Wear (fretting corrosion) and ceramic head fractures are 2 poorly understood concerns related to use, and the role of adapter sleeves has not been defined. In this registry study we assessed the influence of sleeve interposition on re-revision rates in revision hip arthroplasties with retained stems. Confounding factors (demographics, implant-related features) and failures were also analyzed. Patients and methods - We conducted a registry study on 1,127 revised implants (retained trunnion and head exchange). In 26% of implants an adapter sleeve was interposed; in 74% no adapter sleeve was implanted. Demographic and implant-related fe

SUBMITTER: Affatato S 

PROVIDER: S-EPMC6746291 | biostudies-literature | 2019 Oct

REPOSITORIES: biostudies-literature