Comparison of VITEK® 2, three different gradient strip tests and broth microdilution for detecting vanB-positive Enterococcus faecium isolates with low vancomycin MICs.


ABSTRACT:

Objectives

In 2018, EUCAST issued a warning regarding unreliable results of gradient strip tests for confirming vancomycin resistance in enterococci. We compared the performance of various diagnostic standard and confirmatory tests to identify and determine vanB-type vancomycin resistance.

Methods

We analysed a collection of vanB-positive Enterococcus faecium isolates (n = 68) with low vancomycin MICs and compared the performance of VITEK® 2 (bioMérieux), broth microdilution and three gradient strip tests from different providers (Oxoid, Liofilchem and bioMérieux). For the latter we compared the standard procedure with a protocol with increased inoculum, a rich agar medium and a longer incubation time ('macromethod').

Results

The sensitivity of VITEK® 2 was 81% compar