Simethicone is retained in endoscopes despite reprocessing: impact of its use on working channel fluid retention and adenosine triphosphate bioluminescence values (with video).


ABSTRACT:

Background and aims

Studies from our group and others demonstrate residual fluid in 42% to 95% of endoscope working channels despite high-level disinfection and drying. Additionally, persistent simethicone has been reported in endoscope channels despite reprocessing.

Methods

Endoscopy was performed by using water or varied simethicone concentrations (0.5%, 1%, 3%) for flushing. After high-level disinfection/drying, we inspected endoscope working channels for retained fluid by using the SteriCam borescope. Working channel rinsates were evaluated for adenosine triphosphate (ATP) bioluminescence. Fourier transform infrared spectroscopy was performed on fluid droplets gathered from a colonoscope in which low-concentration simethicone was used.

Results

Use of medium/high c