Thin Magnetically Permeable Targets for Inductive Sensing: Application to Limb Prosthetics.


ABSTRACT: The purpose of this research was to create a thin ferrous polymer composite to be used as a target for inductive sensing in limb prosthetics. Inductive sensors are used to monitor limb-to-socket distance in prosthetic sockets, which reflects socket fit. A styrene-ethylene-ethylene/propylene-styrene (SEEPS) polymer was mixed with iron powder at three concentrations (75, 77, 85 wt%), and thin disk-shaped samples were fabricated (0.50, 0,75, 1.00 mm thickness). For 85 wt% samples of 0.50 mm thickness, which proved the best combination of high signal strength and low target volume, inductive sensor sensitivity ranged from 3.2E5 counts/mm at 0.00-1.00 mm distances to 7.2E4 counts/mm at 4.00-5.00 mm distances. The application of compressive stress (up to 425 kPa) introduced an absolute measureme