Comparing mapping and direct hyperspectral imaging in stand-off Raman spectroscopy for remote material identification.
Ontology highlight
ABSTRACT: Stand-off Raman spectroscopy offers a highly selective technique to probe unknown substances from a safe distance. Often, it is necessary to scan large areas of interest. This can be done by pointwise imaging (PI), that is, spectra are sequentially acquired from an array of points over the region of interest (point-by-point mapping). Alternatively, in this paper a direct hyperspectral Raman imager is presented, where a defocused laser beam illuminates a wide area of the sample and the Raman scattered light is collected from the whole field of view (FOV) at once as a spectral snapshot filtered by a liquid crystal tunable filter to select a specific Raman shift. Both techniques are compared in terms of achievable FOV, spectral resolution, signal-to-noise performance, and time consumption dur
SUBMITTER: Gasser C
PROVIDER: S-EPMC6774338 | biostudies-literature | 2019 Jul
REPOSITORIES: biostudies-literature
ACCESS DATA