Ontology highlight
ABSTRACT: Background
In the ongoing efforts to reduce cardiac perfusion dose (injected radioactivity) for conventional SPECT/CT systems, we performed a human observer study to confirm our clinical model observer findings that iterative reconstruction employing OSEM (ordered-subset expectation-maximization) at 25% of the full dose (quarter-dose) has a similar performance for detection of hybrid cardiac perfusion defects as FBP at full dose.Methods
One hundred and sixty-six patients, who underwent routine rest-stress Tc-99m sestamibi cardiac perfusion SPECT/CT imaging and clinically read as normally perfused, were included in the study. Ground truth was established by the normal read and the insertion of hybrid defects. In addition to the reconstruction of the 25% of full-dose data usi
SUBMITTER: Pretorius PH
PROVIDER: S-EPMC6842418 | biostudies-literature | 2021 Apr
REPOSITORIES: biostudies-literature