Multicenter CT phantoms public dataset for radiomics reproducibility tests.


ABSTRACT:

Purpose

The aim of this paper is to describe a public, open-access, computed tomography (CT) phantom image set acquired at three centers and collected especially for radiomics reproducibility research. The dataset is useful to test radiomic features reproducibility with respect to various parameters, such as acquisition settings, scanners, and reconstruction algorithms.

Acquisition and validation methods

Three phantoms were scanned in three independent institutions. Images of the following phantoms were acquired: Catphan 700 and COPDGene Phantom II (Phantom Laboratory, Greenwich, NY, USA), and the Triple modality 3D Abdominal Phantom (CIRS, Norfolk, VA, USA). Data were collected at three Dutch medical centers: MAASTRO Clinic (Maastricht, NL), Radboud University Medical Cente