ABSTRACT: Introduction
Heparin-induced thrombocytopenia (HIT) is a potentially life-threatening adverse reaction of heparin. Laboratory evaluation of HIT is often not available within a reasonable time. We evaluated the HemosIL® HIT-Ab(PF4-H) (Instrumentation Laboratory), a rapid, on-demand, fully automated, latex immunoturbidimetric assay (LIA).Materials and methods
Following determination of the LIA's reference interval and cutoff values, a multicenter study was conducted between March 2013 and June 2015. Plasma samples of HIT-suspected patients (n = 632) were collected and evaluated by LIA on the ACL TOP® Family systems (Instrumentation Laboratory), enzyme-linked immunosorbent assays (EIA), and serotonin release assay (SRA). Patient characteristics
SUBMITTER: Refaai MA
PROVIDER: S-EPMC6850659 | biostudies-literature | 2019 Aug
REPOSITORIES: biostudies-literature