Comparability of three intraocular pressure measurement: iCare pro rebound, non-contact and Goldmann applanation tonometry in different IOP group.


ABSTRACT: BACKGROUND:Measurement of intraocular pressure (IOP) is essential for glaucoma patients. Many factors such as central corneal thickness (CCT) can affect the accuracy of IOP measurement. The purpose of this study was to evaluate the agreement of IOP measured by non-contact tonometer (NCT), iCare pro rebound tonometer (iCare), and Goldmann applanation tonometer (GAT) in different IOP group. METHODS:This was a Hospital-based cross-sectional study. Two hundred subjects were enrolled in this study. All subjects underwent IOP measurement using an NCT-iCare-GAT sequence. Bland-Altman, Pearson correlation and intraclass correlation analysis were performed using SPSS 17.0 software. The influence of CCT on each IOP measurement methods was evaluated by linear regression analysis. RESULTS:The mean dif