Repeatability of Optical Coherence Tomography Angiography in Uveitic Eyes.


ABSTRACT:

Purpose

To investigate the intravisit repeatability of optical coherence tomography angiography (OCTA) in a cohort of uveitis patients.

Methods

One hundred ten patients were imaged twice per eye, per visit, using the Zeiss Cirrus HD-OCT Model 5000 device. To calculate choriocapillaris flow void area (CC FV) 6 × 6-mm images were used, and 3 × 3-mm images were used to calculate vessel density (VD) and the foveal avascular zone area (FAZ) of the superficial capillary plexus (SCP) and deep capillary plexus (DCP). Repeatability was measured using Bland-Altman analyses and intraclass correlation coefficients (ICC) with associated coefficient of variation (CV).

Results

The level of intravisit repeatability differed across indices ranging from moderate to excellent. CC FV had