Unknown

Dataset Information

Missing head and color banding in low-count SPECT reconstructions.


ABSTRACT: Due to low counts in an (111)In single photon emission computed tomography (SPECT) scan, a large part of the head was missing in the reconstructed images on Philips Extended Brilliance Workspace (EBW) and IntelliSpace Portal (ISP) workstations. This problem occurred for the ordered subsets expectation maximization (OSEM) algorithm with and without resolution recovery (Astonish), but not for filtered backprojection (FBP) or maximum likelihood expectation maximization (MLEM). There were also underflow problems because the images are stored as integers resulting in a loss of intensity resolution and color banding.Philips EBW2.0 and ISP5.02 workstations upscale low-count images, but the result is not always optimal, for example, in the case of low counts in one part and more counts in another

SUBMITTER: de Nijs R 

PROVIDER: S-EPMC6890917 | biostudies-literature | 2014 Dec

REPOSITORIES: biostudies-literature