Differences in attenuation pattern in myocardial SPECT between CZT and conventional gamma cameras.


ABSTRACT:

Background

In myocardial perfusion imaging (MPI), single-photon emission tomography (SPECT) soft-tissue attenuation by the abdomen, breasts, and lateral chest wall may create artifacts that mimic true perfusion defects. This may cause misdiagnosis of myocardial perfusion. The aim of the present study was to compare the localization, extent, and depth of attenuation artifacts in MPI SPECT for a multi-pinhole cadmium zinc telluride (CZT) camera vs a conventional gamma camera.

Methods

Phantom and patient measurements were performed using a CZT camera (GE NM 530c) and a conventional gamma camera (GE Ventri). All images were attenuation corrected with externally acquired low-dose computed tomography. The localization, extent, and depth of the attenuation artifact were quantified