ABSTRACT: Background
Detecting asymptomatic and undiagnosed atrial fibrillation (AF) is increasingly important. Recently, we developed a wristwatch-based pulse wave monitor (PWM; Seiko Epson, Japan) capable of long-term recording, with an automatic diagnosis algorithm that uses frequency-based pulse wave analysis. The aim of this study was to evaluate the validity of continuous pulse wave monitoring for detection of AF.Methods
During the electrophysiological study (EPS) in patients with AF, simultaneous pulse wave monitoring and Holter electrocardiograms (ECG) were recorded (n = 136, mean age 62.7 ± 10.9 years). The diagnostic accuracy of the PWM for AF was compared to the Holter ECG diagnosis. Standard performance metrics (sensitivity [Se], specificity [Sp], positive predictive valu
SUBMITTER: Kashiwa A
PROVIDER: S-EPMC6931792 | biostudies-literature | 2019 Mar