Validity and reliability evidence of a point of care assessment of salivary cortisol and ?-amylase: a pre-registered study.


ABSTRACT: Purpose:The iPro Cube is a small portable point-of-care device designed to analyse salivary markers of stress in a user-friendly manner (e.g., fast, convenient). Our aim was to test the reliability and validity of the iPro Cube to measure salivary cortisol and ?-amylase as compared to the common laboratory standard method (ELISA immunoassay) prior to and after moderate intensity exercise. Methods:The study was a repeated measures, pre-registered design, and statistical framework that incorporated prior knowledge directly into the estimation process. Twenty-nine individuals (age = 27.4 ± 6.6 y; body-mass = 70.8 ± 11.3 kg; height = 1.74 ± 0.92 m; 18 males) completed a single PWC75%HRmax, with repeated measures of salivary cortisol and -amylase pre, immediately post, and 30 min post-exercise. Results:Correlation between the iPro Cube and laboratory-based assessments of salivary cortisol was moderate-to-large (0.53 > r < 0.81) across all three testing points. In contrast, correlation between the iPro Cube and laboratory-based assessments of -amylase was small-to-moderate (0.25 > r < 0.46). We found a large correlation between duplicate samples of iPro Cube cortisol assessment (0.75 > r < 0.82), and a moderate-to-large correlation for -amylase (0.51> r < 0.77). Conclusions:The iPro Cube is capable of taking measures of salivary cortisol that are moderately correlated to values obtained via ELISA immunoassay, however the unit underestimates salivary cortisol and overestimates salivary -amylase at rest and post-moderate intensity exercise. It is recommended that researchers continue using standard laboratory techniques to assess these salivary stress markers.

SUBMITTER: Ducker KJ 

PROVIDER: S-EPMC6954686 | biostudies-literature | 2020

REPOSITORIES: biostudies-literature

altmetric image

Publications

Validity and reliability evidence of a point of care assessment of salivary cortisol and <i>α</i>-amylase: a pre-registered study.

Ducker Kagan J KJ   Lines Robin L J RLJ   Chapman Michael T MT   Peeling Peter P   McKay Alannah K A AKA   Gucciardi Daniel F DF  

PeerJ 20200108


<h4>Purpose</h4>The iPro Cube is a small portable point-of-care device designed to analyse salivary markers of stress in a user-friendly manner (e.g., fast, convenient). Our aim was to test the reliability and validity of the iPro Cube to measure salivary cortisol and <i>α</i>-amylase as compared to the common laboratory standard method (ELISA immunoassay) prior to and after moderate intensity exercise.<h4>Methods</h4>The study was a repeated measures, pre-registered design, and statistical fram  ...[more]

Similar Datasets

| S-EPMC8756664 | biostudies-literature
| S-EPMC9473392 | biostudies-literature
| S-EPMC9012170 | biostudies-literature
| S-EPMC5946548 | biostudies-literature
| S-EPMC3629192 | biostudies-literature
| S-EPMC6719785 | biostudies-literature
| S-EPMC11661652 | biostudies-literature
| S-EPMC3408464 | biostudies-literature
| S-EPMC8360508 | biostudies-literature
| S-EPMC7554812 | biostudies-literature