Ontology highlight
ABSTRACT: Aim
To compare clinical outcomes between Ahmed glaucoma implant (AGV-S2 and FP7 models) and Baerveldt-350 glaucoma implant (BGI).Design
Prospective randomized study.Methods and subjects
Eighty-one participants with glaucoma after ocular surgery or secondary glaucoma with persistent and uncontrolled IOP > 21 mmHg were randomized for placement of BGI or AGV models using a standardized surgical technique. The primary outcome was failure, which was defined as IOP >16 mmHg with glaucoma medication, on 2 consecutive study visits. Secondary outcomes were IOP, medication use, visual acuity, complications, and interventions.Results
At one-year follow up, the mean IOP was 14.76±2.5 mmHg in BGI group and 16.57±3.35 mmHg in AGV group (P=0.015). The mean number of glauco
SUBMITTER: Elbaklish KH
PROVIDER: S-EPMC6955622 | biostudies-literature | 2020