High-throughput radio-TLC analysis.


ABSTRACT:

Introduction

Radio thin layer chromatography (radio-TLC) is commonly used to analyze purity of radiopharmaceuticals or to determine the reaction conversion when optimizing radiosynthesis processes. In applications where there are few radioactive species, radio-TLC is preferred over radio-high-performance liquid chromatography due to its simplicity and relatively quick analysis time. However, with current radio-TLC methods, it remains cumbersome to analyze a large number of samples during reaction optimization. In a couple of studies, Cerenkov luminescence imaging (CLI) has been used for reading radio-TLC plates spotted with a variety of isotopes. We show that this approach can be extended to develop a high-throughput approach for radio-TLC analysis of many samples.

Methods

The high-throughput radio-TLC analysis was carried out by performing parallel development of multiple radioactive samples spotted on a single TLC plate, followed by simultaneous readout of the separated samples using Cerenkov imaging. Using custom-written MATLAB software, images were processed and regions of interest (ROIs) were drawn to enclose the radioactive regions/spots. For each sample, the proportion of integrated signal in each ROI was computed. Various crude samples of [18F]fallypride, [18F]FET and [177Lu]Lu-PSMA-617 were prepared for demonstration of this new method.

Results

Benefiting from a parallel developing process and high resolution of CLI-based readout, total analysis time for eight [18F]fallypride samples was 7.5 min (2.5 min for parallel developing, 5 min for parallel readout), which was significantly shorter than the 48 min needed using conventional approaches (24 min for sequential developing, 24 min for sequential readout on a radio-TLC scanner). The greater separation resolution of CLI enabled the discovery of a low-abundance side product from a crude [18F]FET sample that was not discernable using the radio-TLC scanner. Using the CLI-based readout method, we also observed that high labeling efficiency (99%) of [177Lu]Lu-PSMA-617 can be achieved in just 10 min, rather than the typical 30 min timeframe used.

Conclusions

Cerenkov imaging in combination with parallel developing of multiple samples on a single TLC plate proved to be a practical method for rapid, high-throughput radio-TLC analysis.

SUBMITTER: Wang J 

PROVIDER: S-EPMC6956702 | biostudies-literature | 2020 Mar - Apr

REPOSITORIES: biostudies-literature

altmetric image

Publications

High-throughput radio-TLC analysis.

Wang Jia J   Rios Alejandra A   Lisova Ksenia K   Slavik Roger R   Chatziioannou Arion F AF   van Dam R Michael RM  

Nuclear medicine and biology 20191217


<h4>Introduction</h4>Radio thin layer chromatography (radio-TLC) is commonly used to analyze purity of radiopharmaceuticals or to determine the reaction conversion when optimizing radiosynthesis processes. In applications where there are few radioactive species, radio-TLC is preferred over radio-high-performance liquid chromatography due to its simplicity and relatively quick analysis time. However, with current radio-TLC methods, it remains cumbersome to analyze a large number of samples during  ...[more]

Similar Datasets

| S-EPMC3696418 | biostudies-literature
| S-EPMC3075913 | biostudies-literature
| S-EPMC4363487 | biostudies-other
| S-EPMC3128206 | biostudies-literature
| S-EPMC7319559 | biostudies-literature
| S-EPMC5805751 | biostudies-literature
| S-EPMC3957073 | biostudies-literature
| S-EPMC4538797 | biostudies-literature
| S-EPMC3183799 | biostudies-other