Evaluation of a laboratory-developed test for simultaneous detection of norovirus and rotavirus by real-time RT-PCR on the Panther Fusion® system.


ABSTRACT: The Hologic Panther Fusion® Open Access™ functionality allows implementation of laboratory-developed tests (LDTs), with fully automated sample extraction, real-time PCR, and result interpretation. We report the development and validation of a multiplex LDT for norovirus G1, norovirus G2, and rotavirus from stool samples on this system. The LDT was optimized for primer and probe sequences, salt concentration, and PCR annealing temperature. Reproducibility of the PCR and extraction process was assessed. Performance of the multiplex LDT assay was evaluated with external quality assessment (EQA) samples and compared to a commercial multiplex assay (Allplex™ GI-Virus Assay, Seegene) in clinical samples. Salt concentrations and annealing/extension temperature were optimized to 4 mM MgCl2, 70 mM