Impact of detector selections on inter-institutional variability of flattening filter-free beam data for TrueBeam™ linear accelerators.
Ontology highlight
ABSTRACT: This study evaluates the type of detector influencing the inter-institutional variability in flattening filter-free (FFF) beam-specific parameters for TrueBeam™ linear accelerators (Varian Medical Systems,Palo Alto, CA, USA). Twenty-four beam data sets, including the percent depth dose (PDD), off-center ratio (OCR), and output factor (OPF) for modeling within the Eclipse (Varian Medical Systems) treatment planning system, were collected from 19 institutions. Although many institutions collected the data using CC13 (IBA Dosimetry, Schwarzenbruck, Germany) or PTW31010 semiflex (PTW Freiburg, Freiburg, Germany) ionization chambers, some institutions used diode detectors, diamond detectors, and ionization chambers with smaller cavities. The OCR data included penumbra width, full width at half