Small-field dosimetry of TrueBeamTM flattened and flattening filter-free beams: A multi-institutional analysis.


ABSTRACT:

Purpose

Detector-dependent interinstitutional variations of the beam data may lead to uncertainties of the delivered dose to patients. Here we evaluated the inter-unit variability of the flattened and flattening filter-free (FFF) beam data of multiple TrueBeam (Varian Medical Systems) linear accelerators focusing on the small-field dosimetry.

Methods

The beam data of 6- and 10-MV photon beams with and without flattening filter measured for modeling of an iPLAN treatment planning system (BrainLAB) were collected from 12 institutions - ten HD120 Multileaf Collimator (MLC) and two Millennium120 MLC. Percent-depth dose (PDD), off-center ratio (OCR), and detector output factors (OFdet ) measured with different detectors were evaluated. To investigate the detector-associated effects, we evaluated the inter-unit variations of the OFdet before and after having applied the output correction factors provided by the International Atomic Energy Agency (IAEA) Technical Reports Series no. 483.

Results

PDD measured with a field size of 5 × 5 mm2 showed that the data measured using an ionization chamber had variations exceeding 1% from the median values. The maximum difference from median value was 2.87% for 10 MV photon beam. The maximum variations of the penumbra width for OCR with 10 × 10 mm2 field size were 0.97 mm. The OFdet showed large variations exceeding 15% for a field size of 5 × 5 mm2 . When the output correction factors were applied to the OFdet , the variations were greatly reduced. The relative difference of almost all field output factors were within ± 5% from the median field output factors.

Conclusion

In this study, the inter-unit variability of small-field dosimetry was evaluated for TrueBeam linear accelerators. The variations were large at a field size of 5 × 5 mm2 , and most occurred in a detector-dependent manner.

SUBMITTER: Akino Y 

PROVIDER: S-EPMC6964782 | biostudies-literature | 2020 Jan

REPOSITORIES: biostudies-literature

altmetric image

Publications

Small-field dosimetry of TrueBeam<sup>TM</sup> flattened and flattening filter-free beams: A multi-institutional analysis.

Akino Yuichi Y   Mizuno Hirokazu H   Isono Masaru M   Tanaka Yoshihiro Y   Masai Norihisa N   Yamamoto Toshijiro T  

Journal of applied clinical medical physics 20191209 1


<h4>Purpose</h4>Detector-dependent interinstitutional variations of the beam data may lead to uncertainties of the delivered dose to patients. Here we evaluated the inter-unit variability of the flattened and flattening filter-free (FFF) beam data of multiple TrueBeam (Varian Medical Systems) linear accelerators focusing on the small-field dosimetry.<h4>Methods</h4>The beam data of 6- and 10-MV photon beams with and without flattening filter measured for modeling of an iPLAN treatment planning s  ...[more]