Technical Considerations and Confounders for Urine CXCL10 Chemokine Measurement.


ABSTRACT:

Background

The urine C-X-C motif chemokine 10 (CXCL10) is a promising screening biomarker for renal allograft rejection. The aim of the study was to investigate important technical and biological aspects as well as potential confounders when measuring urine CXCL10.

Methods

We analyzed 595 urine samples from 117 patients, who participated in a randomized controlled trial investigating the clinical utility of urine CXCL10 monitoring for posttransplant management. Urine CXCL10 was measured by an immunoassay using electrochemiluminescence.

Results

Intraassay coefficient of variation was 2.5%, and interassay coefficient of variation was 10%. Urine CXCL10 remained stable (ie, <10% degradation) for 8 hours at 25°C or 37°C and for 3 days at 4°C. CXCL10 concentrations [pg/mL]