ABSTRACT: Background
Various instruments have been developed to measure aligning prism, the prism that eliminates a fixation disparity (associated heterophoria). This includes the established Mallett near vision unit and recently developed Thomson Vision Toolbox on the iPad. With no previous research investigating the agreement between these instruments, practitioners may question if they can be used interchangeably.Methods
80 participants underwent near vision testing with the Mallett unit and iPad fixation disparity test. Data were analysed in four ways to investigate the agreement of the two instruments.Results
Many participants reported no fixation disparity (horizontally 46.25%, vertically 82.5%), or non-significant aligning prism (horizontally 70%, vertically 97.5%), on