Quality control of multiplex antibody detection in samples from large-scale surveys: the example of malaria in Haiti.
Ontology highlight
ABSTRACT: Measuring antimalarial antibodies can estimate transmission in a population. To compare outputs, standardized laboratory testing is required. Here we describe the in-country establishment and quality control (QC) of a multiplex bead assay (MBA) for three sero-surveys in Haiti. Total IgG data against 21 antigens were collected for 32,758 participants. Titration curves of hyperimmune sera were included on assay plates, assay signals underwent 5-parameter regression, and inspection of the median and interquartile range (IQR) for the y-inflection point was used to determine assay precision. The medians and IQRs were similar for Surveys 1 and 2 for most antigens, while the IQRs increased for some antigens in Survey 3. Levey-Jennings charts for selected antigens provided a pass/fail criterion fo
SUBMITTER: van den Hoogen LL
PROVIDER: S-EPMC6981173 | biostudies-literature | 2020 Jan
REPOSITORIES: biostudies-literature
ACCESS DATA