Trabecular bone imaging using a 3D adiabatic inversion recovery prepared ultrashort TE Cones sequence at 3T.


ABSTRACT:

Purpose

To investigate direct imaging of trabecular bone using a 3D adiabatic inversion recovery prepared ultrashort TE cones (3D IR-UTE-Cones) sequence.

Methods

The proposed 3D IR-UTE-Cones sequence used a broadband adiabatic inversion pulse together with a short TR/TI combination to suppress signals from long T2 tissues such as muscle and marrow fat, followed by multispoke UTE acquisition to detect signal from short T2 water components in trabecular bone. The feasibility of this technique for robust suppression of long T2 tissues was first demonstrated through numerical simulations. The proposed IR-UTE-Cones sequence was applied to a hip agarose bone phantom and to 6 healthy volunteers for morphologic and quantitative T2∗ and proton density