Robotic CT-guided out-of-plane needle insertion: comparison of angle accuracy with manual insertion in phantom and measurement of distance accuracy in animals.


ABSTRACT:

Objectives

To evaluate the accuracy of robotic CT-guided out-of-plane needle insertion in phantom and animal experiments.

Methods

A robotic system (Zerobot), developed at our institution, was used for needle insertion. In the phantom experiment, 12 robotic needle insertions into a phantom at various angles in the XY and YZ planes were performed, and the same insertions were manually performed freehand, as well as guided by a smartphone application (SmartPuncture). Angle errors were compared between the robotic and smartphone-guided manual insertions using Student's t test. In the animal experiment, 6 robotic out-of-plane needle insertions toward targets of 1.0 mm in diameter placed in the kidneys and hip muscles of swine were performed, each with and without adjustment of ne