ABSTRACT: Objective
Accuracy of 18 current-generation blood glucose monitoring systems (BGMS) available in Europe was evaluated applying criteria adapted from EN ISO 15197:2015 with one reagent system lot. BGMS were selected based on market research data.Research design and methods
The BGMS ABRA, Accu-Chek Guide, AURUM, CareSens Dual, CERA-CHEK 1CODE, ContourNext One, eBsensor, FreeStyle Freedom Lite, GL50 evo, GlucoCheck GOLD, GlucoMen areo 2K, GluNEO, MyStar DoseCoach, OneTouch Verio Flex, Pic GlucoTest, Rightest GM700S, TRUEyou, and WaveSense JAZZ Wireless were tested using capillary blood from 100 different subjects and assessing the percentage of results within ±15 mg/dL (0.83 mmol/L) or 15% of comparison method results for BG concentrations below or above 100 mg/dL (5.55 mmol/L