ABSTRACT: Rationale
Strontium isotopes are valuable markers of provenance in a range of disciplines. Limited amounts of Sr in low-mass samples such as insects mean that conventional Sr isotope analysis precludes their use for geographic origins in many ecological studies or in applications such as biosecurity. Here we test the viability of using inductively coupled plasma tandem mass spectrometry (ICP-MS/MS) with N2 O as a reaction gas for accurately determining Sr isotopes in insects with Sr < 100 ng.Methods
Strontium isotopes were determined in solution mode using ICP-MS/MS with 0.14 L/min N2 O as a reaction gas to convert Sr+ into SrO+ for in-line separation of 87 Sr from 87 Rb. The Sr isotope reference standards NI