Unknown

Dataset Information

Endoscopic optical coherence tomography angiography using a forward imaging piezo scanner probe.


ABSTRACT: A forward imaging endoscope for optical coherence tomography angiography (OCTA) featuring a piezoelectric fiber scanner is presented. Imaging is performed with an optical coherence tomography (OCT) system incorporating an akinetic light source with a center wavelength of 1300 nm, bandwidth of 90 nm and A-line rate of 173 kHz. The endoscope operates in contact mode to avoid motion artifacts, in particular, beneficial for OCTA measurements, and achieves a transversal resolution of 12 μm in air at a rigid probe size of 4 mm in diameter and 11.3 mm in length. A spiral scan pattern is generated at a scanning frequency of 360 Hz to sample a maximum field of view of 1.3 mm. OCT images of a human finger as well as visualization of microvasculature of the human palm are presented both in two and th

SUBMITTER: Wurster LM 

PROVIDER: S-EPMC7065608 | biostudies-literature | 2019 Apr

REPOSITORIES: biostudies-literature