Ontology highlight
ABSTRACT: Objectives
When patients with acute respiratory distress syndrome are moved out of an intensive care unit, the ventilator often requires changing. This procedure suppresses positive end expiratory pressure and promotes lung derecruitment. Clamping the endotracheal tube may prevent this from occurring. Whether or not such clamping maintains positive end-expiratory pressure has never been investigated. We designed a bench study to explore this further.How the study was done
We used the Elysee 350 ventilator in 'volume controlled' mode with a positive end-expiratory pressure of 15 cmH2O, connected to an endotracheal tube with an 8 mm internal diameter inserted into a lung model with 40 ml/cmH2O compliance and 10 cmH2O/L/s resistance. We measured airway pressure and flow betwee
SUBMITTER: Turbil E
PROVIDER: S-EPMC7065807 | biostudies-literature | 2020
REPOSITORIES: biostudies-literature