Assay of Phospholipase D Activity by an Amperometric Choline Oxidase Biosensor.
Ontology highlight
ABSTRACT: A novel electrochemical method to assay phospholipase D (PLD) activity is proposed based on the employment of a choline biosensor realized by immobilizing choline oxidase through co-crosslinking on an overoxidized polypyrrole film previously deposited on a platinum electrode. To perform the assay, an aliquot of a PLD standard solution is typically added to borate buffer containing phosphatidylcholine at a certain concentration and the oxidation current of hydrogen peroxide is then measured at the rotating modified electrode by applying a detection potential of + 0.7 V vs. SCE. Various experimental parameters influencing the assay were studied and optimized. The employment of 0.75% (v/v) Triton X-100, 0.2 mM calcium chloride, 5 mM phosphatidylcholine, and borate buffer at pH 8.0, ionic
SUBMITTER: Ciriello R
PROVIDER: S-EPMC7085753 | biostudies-literature | 2020 Feb
REPOSITORIES: biostudies-literature
ACCESS DATA