ABSTRACT: Purpose
To compare filter tilt and filter jumping during Option inferior vena cava (IVC) filter deployment with 3 different wires techniques using a 3-dimensional (3D) printing vena cava phantom.Materials and methods
An IVC 3D printed vena cava phantom was made from a healthy young male's computed tomographic data. Option IVC filters were deployed with 3 different wires: i) original push wire, ii) hydrophilic stiff wire, and iii) bent stiff wire. Right internal jugular and right femoral access were used 5 times with each wire. Filter tilt angle, tilt ratio, jumping, and tip abutment to the IVC wall were analyzed.Results
The transfemoral approach with original push wire had significantly higher tilt angle than did the transjugular approach (6.1˚ ± 1.9 vs. 3.5˚ ± 1.3,