Simplified pulse wave velocity measurement in children: Is the pOpmetre valid?
Ontology highlight
ABSTRACT: In population exposed to cardiovascular risk, aortic stiffness is an important marker which is assessed by carotid-to-femoral pulse wave velocity (PWV). In childhood, the validated applanation tonometer SphygmoCor® can be used to measure PWV, but is limited in routine practice by the child's cooperation and operator's experience. An alternative device, the pOpmètre® is validated in adults and rapidly measures finger-to-toe PWV using 2 oxymeter-like sensors. The aim of this study is to validate the pOpmètre® device in children aged between 4 and 8 years. We compared simultaneous PWV measurements of the two devices, SphygmoCor® and pOpmètre®, in a training group, using the Bland-Altman method. Then we proposed an algorithm to correct pOpmètre® PWV (PWVpop). Finally, we validated this new alg