Unknown

Dataset Information

Data-Driven Respiratory Gating Outperforms Device-Based Gating for Clinical 18F-FDG PET/CT.


ABSTRACT: A data-driven method for respiratory gating in PET has recently been commercially developed. We sought to compare the performance of the algorithm with an external, device-based system for oncologic 18F-FDG PET/CT imaging. Methods: In total, 144 whole-body 18F-FDG PET/CT examinations were acquired, with a respiratory gating waveform recorded by an external, device-based respiratory gating system. In each examination, 2 of the bed positions covering the liver and lung bases were acquired with a duration of 6 min. Quiescent-period gating retaining approximately 50% of coincidences was then able to produce images with an effective duration of 3 min for these 2 bed positions, matching the other bed positions. For each examination, 4 reconstructions were performed a

SUBMITTER: Walker MD 

PROVIDER: S-EPMC7116324 | biostudies-literature | 2020 Nov

REPOSITORIES: biostudies-literature