Excess pressure as an analogue of blood flow velocity.


ABSTRACT:

Introduction

Derivation of blood flow velocity from a blood pressure waveform is a novel technique, which could have potential clinical importance. Excess pressure, calculated from the blood pressure waveform via the reservoir-excess pressure model, is purported to be an analogue of blood flow velocity but this has never been examined in detail, which was the aim of this study.

Methods

Intra-arterial blood pressure was measured sequentially at the brachial and radial arteries via fluid-filled catheter simultaneously with blood flow velocity waveforms recorded via Doppler ultrasound on the contralateral arm (n = 98, aged 61 ± 10 years, 72% men). Excess pressure was derived from intra-arterial blood pressure waveforms using pressure-only reservoir-excess pressure analysis.