Atrial signal amplitude predicts atrial high-rate episodes in implantable cardioverter defibrillator patients: Insights from a large database of remote monitoring transmissions.


ABSTRACT:

Background

Parameters measured during implantable cardioverter defibrillator (ICD) implant also depend on bioelectrical properties of the myocardium. We aimed to explore their potential association with clinical outcomes in patients with single/dual-chamber ICD and cardiac resynchronization therapy defibrillator (CRT-D).

Methods

In the framework of the Home Monitoring Expert Alliance, baseline electrical parameters for all implanted leads were compared by the occurrence of all-cause mortality, adjudicated ventricular arrhythmia (VA), and atrial high-rate episode lasting ≥24 hours (24 h AHRE).

Results

In a cohort of 2976 patients (58.1% ICD) with a median follow-up of 25 months, event rates were 3.1/100 patient-years for all-cause mortality, 18.1/100 patient-years for