Unknown

Dataset Information

In Vitro Durability and Stability Testing of a Novel Polymeric Transcatheter Aortic Valve.


ABSTRACT: Transcatheter aortic valve replacement (TAVR) has emerged as an effective therapy for the unmet clinical need of inoperable patients with severe aortic stenosis (AS). Current clinically used tissue TAVR valves suffer from limited durability that hampers TAVR's rapid expansion to younger, lower risk patients. Polymeric TAVR valves optimized for hemodynamic performance, hemocompatibility, extended durability, and resistance to calcific degeneration offer a viable solution to this challenge. We present extensive in vitro durability and stability testing of a novel polymeric TAVR valve (PolyNova valve) using 1) accelerated wear testing (AWT, ISO 5840); 2) calcification susceptibility (in the AWT)-compared with clinically used tissue valves; and 3) extended crimping stability (valves crimped to

SUBMITTER: Rotman OM 

PROVIDER: S-EPMC7136031 | biostudies-literature | 2020 Feb

REPOSITORIES: biostudies-literature