Validation of an LC-MS/MS-based dilute-and-shoot approach for the quantification of > 500 mycotoxins and other secondary metabolites in food crops: challenges and solutions.
Ontology highlight
ABSTRACT: This paper describes the validation of an LC-MS/MS-based method for the quantification of > 500 secondary microbial metabolites. Analytical performance parameters have been determined for seven food matrices using seven individual samples per matrix for spiking. Apparent recoveries ranged from 70 to 120% for 53-83% of all investigated analytes (depending on the matrix). This number increased to 84-94% if the recovery of extraction was considered. The comparison of the fraction of analytes for which the precision criterion of RSD ≤ 20% under repeatability conditions (for 7 replicates derived from different individual samples) and intermediate precision conditions (for 7 technical replicates from one sample), respectively, was met (85-97% vs. 93-94%) highlights the contribution of relative m
SUBMITTER: Sulyok M
PROVIDER: S-EPMC7136310 | biostudies-literature | 2020 Apr
REPOSITORIES: biostudies-literature