Comparison of feature tracking, fast-SENC, and myocardial tagging for global and segmental left ventricular strain.
Ontology highlight
ABSTRACT: AIMS:A multitude of cardiac magnetic resonance (CMR) techniques are used for myocardial strain assessment; however, studies comparing them are limited. We sought to compare global longitudinal (GLS), circumferential (GCS), segmental longitudinal (SLS), and segmental circumferential (SCS) strain values, as well as reproducibility between CMR feature tracking (FT), tagging (TAG), and fast-strain-encoded (fast-SENC) CMR techniques. METHODS AND RESULTS:Eighteen subjects (11 healthy volunteers and seven patients with heart failure) underwent two CMR scans (1.5T, Philips) with identical parameters. Global and segmental strain values were measured using FT (Medis), TAG (Medviso), and fast-SENC (Myocardial Solutions). Friedman's test, linear regression, Pearson's correlation coefficient, and Bland
SUBMITTER: Bucius P
PROVIDER: S-EPMC7160507 | biostudies-literature | 2020 Apr
REPOSITORIES: biostudies-literature