Feasibility and accuracy of a robotic guidance system for navigated spine surgery in a hybrid operating room: a cadaver study.


ABSTRACT: The combination of navigation and robotics in spine surgery has the potential to accurately identify and maintain bone entry position and planned trajectory. The goal of this study was to examine the feasibility, accuracy and efficacy of a new robot-guided system for semi-automated, minimally invasive, pedicle screw placement. A custom robotic arm was integrated into a hybrid operating room (OR) equipped with an augmented reality surgical navigation system (ARSN). The robot was mounted on the OR-table and used to assist in placing Jamshidi needles in 113 pedicles in four cadavers. The ARSN system was used for planning screw paths and directing the robot. The robot arm autonomously aligned with the planned screw trajectory, and the surgeon inserted the Jamshidi needle into the pedicle. Accuracy measurements were performed on verification cone beam computed tomographies with the planned paths superimposed. To provide a clinical grading according to the Gertzbein scale, pedicle screw diameters were simulated on the placed Jamshidi needles. A technical accuracy at bone entry point of 0.48?±?0.44?mm and 0.68?±?0.58?mm was achieved in the axial and sagittal views, respectively. The corresponding angular errors were 0.94?±?0.83° and 0.87?±?0.82°. The accuracy was statistically superior (p?

SUBMITTER: Burstrom G 

PROVIDER: S-EPMC7200720 | biostudies-literature | 2020 May

REPOSITORIES: biostudies-literature

altmetric image

Publications

Feasibility and accuracy of a robotic guidance system for navigated spine surgery in a hybrid operating room: a cadaver study.

Burström Gustav G   Balicki Marcin M   Patriciu Alexandru A   Kyne Sean S   Popovic Aleksandra A   Holthuizen Ronald R   Homan Robert R   Skulason Halldor H   Persson Oscar O   Edström Erik E   Elmi-Terander Adrian A  

Scientific reports 20200505 1


The combination of navigation and robotics in spine surgery has the potential to accurately identify and maintain bone entry position and planned trajectory. The goal of this study was to examine the feasibility, accuracy and efficacy of a new robot-guided system for semi-automated, minimally invasive, pedicle screw placement. A custom robotic arm was integrated into a hybrid operating room (OR) equipped with an augmented reality surgical navigation system (ARSN). The robot was mounted on the OR  ...[more]

Similar Datasets

| S-EPMC9719204 | biostudies-literature
| S-EPMC11636531 | biostudies-literature
| S-EPMC8132365 | biostudies-literature
| S-EPMC9278478 | biostudies-literature
| PRJEB11484 | ENA
| S-EPMC4868588 | biostudies-other
| S-EPMC8975112 | biostudies-literature
| S-EPMC6381269 | biostudies-literature
| S-EPMC3214254 | biostudies-literature